{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-04-19T06:22:42.779","vulnerabilities":[{"cve":{"id":"CVE-2025-0683","sourceIdentifier":"ics-cert@hq.dhs.gov","published":"2025-01-30T19:15:14.460","lastModified":"2026-04-15T00:35:42.020","vulnStatus":"Deferred","cveTags":[],"descriptions":[{"lang":"en","value":"In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."},{"lang":"es","value":"En su configuración predeterminada, el monitor de pacientes Contec Health CMS8000 transmite datos del paciente en texto plano a una dirección IP pública predefinida cuando el paciente se conecta al monitor. Esto podría provocar una filtración de datos confidenciales del paciente a cualquier dispositivo con esa dirección IP o a un atacante en un escenario de \"máquina en el medio\"."}],"metrics":{"cvssMetricV40":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","cvssData":{"version":"4.0","vectorString":"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":8.2,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","attackRequirements":"PRESENT","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"NONE","vulnAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}}],"cvssMetricV31":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N","baseScore":5.9,"baseSeverity":"MEDIUM","attackVector":"NETWORK","attackComplexity":"HIGH","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"HIGH","integrityImpact":"NONE","availabilityImpact":"NONE"},"exploitabilityScore":2.2,"impactScore":3.6}]},"weaknesses":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","description":[{"lang":"en","value":"CWE-359"}]}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01","source":"ics-cert@hq.dhs.gov"},{"url":"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication","source":"ics-cert@hq.dhs.gov"},{"url":"https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/","source":"af854a3a-2127-422b-91ae-364da2661108"},{"url":"https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor","source":"af854a3a-2127-422b-91ae-364da2661108"},{"url":"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication","source":"af854a3a-2127-422b-91ae-364da2661108"}]}}]}