{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-05-08T02:08:14.020","vulnerabilities":[{"cve":{"id":"CVE-2024-48966","sourceIdentifier":"productsecurity@baxter.com","published":"2024-11-14T22:15:17.727","lastModified":"2026-04-15T00:35:42.020","vulnStatus":"Deferred","cveTags":[],"descriptions":[{"lang":"en","value":"The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance."},{"lang":"es","value":"Las herramientas de software que utiliza el personal de servicio para probar y calibrar el respirador no admiten la autenticación de usuarios. Un atacante con acceso a la PC de servicio donde están instaladas las herramientas podría obtener información de diagnóstico a través de la herramienta de prueba o manipular la configuración del respirador y el software integrado a través de la herramienta de calibración, sin tener que autenticarse en ninguna de las herramientas. Esto podría dar lugar a una divulgación no autorizada de información o tener efectos no deseados en la configuración y el rendimiento del dispositivo."}],"metrics":{"cvssMetricV31":[{"source":"productsecurity@baxter.com","type":"Secondary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","baseScore":10.0,"baseSeverity":"CRITICAL","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"CHANGED","confidentialityImpact":"HIGH","integrityImpact":"HIGH","availabilityImpact":"HIGH"},"exploitabilityScore":3.9,"impactScore":6.0}]},"weaknesses":[{"source":"productsecurity@baxter.com","type":"Secondary","description":[{"lang":"en","value":"CWE-306"}]}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01","source":"productsecurity@baxter.com"}]}}]}