{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-04-17T09:06:48.734","vulnerabilities":[{"cve":{"id":"CVE-2021-33886","sourceIdentifier":"cve@mitre.org","published":"2021-08-25T12:15:17.130","lastModified":"2024-11-21T06:09:43.440","vulnStatus":"Modified","cveTags":[],"descriptions":[{"lang":"en","value":"An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device."},{"lang":"es","value":"Una vulnerabilidad de saneamiento inapropiado de la entrada en B. Braun SpaceCom2 versiones anteriores a 012U000062, permite que un atacante remoto no autenticado obtenga acceso a la línea de comandos a nivel de usuario pasando una cadena externa sin procesar directamente a las declaraciones printf. El atacante debe estar en la misma red que el dispositivo."}],"metrics":{"cvssMetricV31":[{"source":"cve@mitre.org","type":"Secondary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N","baseScore":8.1,"baseSeverity":"HIGH","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"HIGH","integrityImpact":"HIGH","availabilityImpact":"NONE"},"exploitabilityScore":2.8,"impactScore":5.2},{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","baseScore":8.8,"baseSeverity":"HIGH","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"HIGH","integrityImpact":"HIGH","availabilityImpact":"HIGH"},"exploitabilityScore":2.8,"impactScore":5.9}],"cvssMetricV2":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"2.0","vectorString":"AV:A/AC:L/Au:N/C:P/I:P/A:P","baseScore":5.8,"accessVector":"ADJACENT_NETWORK","accessComplexity":"LOW","authentication":"NONE","confidentialityImpact":"PARTIAL","integrityImpact":"PARTIAL","availabilityImpact":"PARTIAL"},"baseSeverity":"MEDIUM","exploitabilityScore":6.5,"impactScore":6.4,"acInsufInfo":false,"obtainAllPrivilege":false,"obtainUserPrivilege":false,"obtainOtherPrivilege":false,"userInteractionRequired":false}]},"weaknesses":[{"source":"nvd@nist.gov","type":"Primary","description":[{"lang":"en","value":"CWE-134"}]}],"configurations":[{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:bbraun:spacecom2:*:*:*:*:*:*:*:*","versionEndExcluding":"012u000062","matchCriteriaId":"C2D8F2AC-5506-405C-857A-EE060C209A89"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:bbraun:infusomat_large_volume_pump_871305u:-:*:*:*:*:*:*:*","matchCriteriaId":"A6C9420B-B015-4079-8B60-EF9C9C058193"},{"vulnerable":false,"criteria":"cpe:2.3:h:bbraun:spacestation_8713142u:-:*:*:*:*:*:*:*","matchCriteriaId":"D6B64426-0B44-4B68-A80A-2BB6D04B3F88"}]}]}],"references":[{"url":"https://www.bbraunusa.com/en.htm","source":"cve@mitre.org","tags":["Broken Link"]},{"url":"https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/","source":"cve@mitre.org","tags":["Exploit","Third Party Advisory"]},{"url":"https://www.bbraunusa.com/en.htm","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Broken Link"]},{"url":"https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Exploit","Third Party Advisory"]}]}}]}