{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-05-11T17:19:55.550","vulnerabilities":[{"cve":{"id":"CVE-2020-27278","sourceIdentifier":"ics-cert@hq.dhs.gov","published":"2021-03-15T22:15:12.873","lastModified":"2024-11-21T05:20:59.140","vulnStatus":"Modified","cveTags":[],"descriptions":[{"lang":"en","value":"In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, hard-coded credentials in the ventilator allow attackers with physical access to obtain admin privileges for the device’s configuration interface."},{"lang":"es","value":"En Hamilton Medical AG, T1-Ventillator versiones 2.2.3 y anteriores, unas credenciales embebidas en el ventilator permiten a atacantes con acceso físico alcanzar privilegios de administrador para la interfaz de configuración del dispositivo"}],"metrics":{"cvssMetricV31":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:N","baseScore":5.2,"baseSeverity":"MEDIUM","attackVector":"PHYSICAL","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"HIGH","integrityImpact":"LOW","availabilityImpact":"NONE"},"exploitabilityScore":0.9,"impactScore":4.2}],"cvssMetricV2":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"2.0","vectorString":"AV:L/AC:L/Au:N/C:P/I:P/A:N","baseScore":3.6,"accessVector":"LOCAL","accessComplexity":"LOW","authentication":"NONE","confidentialityImpact":"PARTIAL","integrityImpact":"PARTIAL","availabilityImpact":"NONE"},"baseSeverity":"LOW","exploitabilityScore":3.9,"impactScore":4.9,"acInsufInfo":false,"obtainAllPrivilege":false,"obtainUserPrivilege":false,"obtainOtherPrivilege":false,"userInteractionRequired":false}]},"weaknesses":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","description":[{"lang":"en","value":"CWE-798"}]}],"configurations":[{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:hamilton-medical:hamilton-t1_firmware:*:*:*:*:*:*:*:*","versionEndIncluding":"2.2.3","matchCriteriaId":"425D6578-CE4D-4417-B92C-12C10C837EC2"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:hamilton-medical:hamilton-t1:-:*:*:*:*:*:*:*","matchCriteriaId":"3F037FC9-EE0F-4328-9BB1-0628DB4B66FA"}]}]}],"references":[{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-047-01","source":"ics-cert@hq.dhs.gov","tags":["Third Party Advisory","US Government Resource"]},{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-047-01","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Third Party Advisory","US Government Resource"]}]}}]}