{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-05-10T08:14:56.028","vulnerabilities":[{"cve":{"id":"CVE-2020-12035","sourceIdentifier":"ics-cert@hq.dhs.gov","published":"2020-06-29T14:15:11.397","lastModified":"2024-11-21T04:59:09.403","vulnStatus":"Modified","cveTags":[],"descriptions":[{"lang":"en","value":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration."},{"lang":"es","value":"Baxter PrismaFlex todas las versiones, PrisMax todas las versiones anteriores a 3.x, el dispositivo PrismaFlex, contiene una contraseña de servicio embebida que proporciona acceso a información biomédica, configuración del dispositivo, configuración de calibración y configuración de red. Esto podría permitir a un atacante modificar la configuración y la calibración del dispositivo"}],"metrics":{"cvssMetricV31":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:L","baseScore":4.9,"baseSeverity":"MEDIUM","attackVector":"PHYSICAL","attackComplexity":"LOW","privilegesRequired":"LOW","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"NONE","integrityImpact":"HIGH","availabilityImpact":"LOW"},"exploitabilityScore":0.7,"impactScore":4.2}],"cvssMetricV2":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"2.0","vectorString":"AV:L/AC:L/Au:N/C:N/I:P/A:P","baseScore":3.6,"accessVector":"LOCAL","accessComplexity":"LOW","authentication":"NONE","confidentialityImpact":"NONE","integrityImpact":"PARTIAL","availabilityImpact":"PARTIAL"},"baseSeverity":"LOW","exploitabilityScore":3.9,"impactScore":4.9,"acInsufInfo":false,"obtainAllPrivilege":false,"obtainUserPrivilege":false,"obtainOtherPrivilege":false,"userInteractionRequired":false}]},"weaknesses":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","description":[{"lang":"en","value":"CWE-287"}]},{"source":"nvd@nist.gov","type":"Primary","description":[{"lang":"en","value":"CWE-798"}]}],"configurations":[{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:baxter:prismaflex_firmware:*:*:*:*:*:*:*:*","matchCriteriaId":"B8874804-02E2-4E74-94C6-F103DCDC854E"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:baxter:prismaflex:-:*:*:*:*:*:*:*","matchCriteriaId":"8E5DB2B2-5D9E-494B-A860-FE2F4BC4C767"}]}]},{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:baxter:prismax_firmware:*:*:*:*:*:*:*:*","versionEndExcluding":"3.0","matchCriteriaId":"8F197DAD-63A3-4598-BBFC-0014707B3A01"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:baxter:prismax:-:*:*:*:*:*:*:*","matchCriteriaId":"A5EB1A30-A514-437D-B976-451CF003955D"}]}]}],"references":[{"url":"https://www.us-cert.gov/ics/advisories/icsma-20-170-01","source":"ics-cert@hq.dhs.gov","tags":["Not Applicable"]},{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02","source":"nvd@nist.gov","tags":["Third Party Advisory"]},{"url":"https://www.us-cert.gov/ics/advisories/icsma-20-170-01","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Not Applicable"]}],"evaluatorComment":"At the time of analysis, information provided did not align with the data (including CVE number) in the advisory."}}]}