{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-08-03T03:52:54.936","vulnerabilities":[{"cve":{"id":"CVE-2018-8868","sourceIdentifier":"ics-cert@hq.dhs.gov","published":"2018-07-03T01:29:01.877","lastModified":"2026-06-17T02:05:33.167","vulnStatus":"Modified","cveTags":[],"descriptions":[{"lang":"en","value":"Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality."},{"lang":"es","value":"Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor en todas sus versiones y 24952 MyCareLink Monitor en todas sus versiones, contienen código de depuración destinado a probar la funcionalidad de las interfaces de comunicación del monitor, incluida la interfaz entre el monitor y el dispositivo cardíaco implantable. Un atacante con acceso físico al dispositivo puede aplicar las otras vulnerabilidades dentro de este advisory para acceder a esta funcionalidad de depuración. Esta funcionalidad de depuración proporciona la capacidad de leer y escribir valores de memoria arbitrarios en dispositivos cardíacos implantables mediante protocolos inalámbricos inductivos o de corto alcance. Un atacante con proximidad física cercana a un dispositivo cardíaco implantable objetivo puede utilizar esta funcionalidad de depuración."}],"affected":[{"source":"ics-cert@hq.dhs.gov","affectedData":[{"vendor":"Medtronic","product":"24950 MyCareLink Monitor","defaultStatus":"unaffected","versions":[{"version":"All versions","status":"affected"}]},{"vendor":"Medtronic","product":"24952 MyCareLink Monitor","defaultStatus":"unaffected","versions":[{"version":"All versions","status":"affected"}]}]}],"metrics":{"cvssMetricV31":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L","baseScore":6.2,"baseSeverity":"MEDIUM","attackVector":"PHYSICAL","attackComplexity":"HIGH","privilegesRequired":"LOW","userInteraction":"NONE","scope":"CHANGED","confidentialityImpact":"HIGH","integrityImpact":"LOW","availabilityImpact":"LOW"},"exploitabilityScore":0.4,"impactScore":5.3}],"cvssMetricV30":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"3.0","vectorString":"CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H","baseScore":6.4,"baseSeverity":"MEDIUM","attackVector":"PHYSICAL","attackComplexity":"HIGH","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"HIGH","integrityImpact":"HIGH","availabilityImpact":"HIGH"},"exploitabilityScore":0.5,"impactScore":5.9}],"cvssMetricV2":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"2.0","vectorString":"AV:L/AC:M/Au:N/C:C/I:C/A:C","baseScore":6.9,"accessVector":"LOCAL","accessComplexity":"MEDIUM","authentication":"NONE","confidentialityImpact":"COMPLETE","integrityImpact":"COMPLETE","availabilityImpact":"COMPLETE"},"baseSeverity":"MEDIUM","exploitabilityScore":3.4,"impactScore":10.0,"acInsufInfo":false,"obtainAllPrivilege":false,"obtainUserPrivilege":false,"obtainOtherPrivilege":false,"userInteractionRequired":false}]},"weaknesses":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","description":[{"lang":"en","value":"CWE-749"}]},{"source":"nvd@nist.gov","type":"Primary","description":[{"lang":"en","value":"NVD-CWE-noinfo"}]}],"configurations":[{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:medtronic:24950_mycarelink_monitor_firmware:-:*:*:*:*:*:*:*","matchCriteriaId":"778EB98D-C77A-4C81-B7AD-848C335B850D"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:medtronic:24950_mycarelink_monitor:-:*:*:*:*:*:*:*","matchCriteriaId":"D8F046CD-32D6-44D1-AB08-D81B3942CB78"}]}]},{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:medtronic:24952_mycarelink_monitor_firmware:-:*:*:*:*:*:*:*","matchCriteriaId":"2C9FE79C-C221-4E6B-A687-74A5D7A85EFA"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:medtronic:24952_mycarelink_monitor:-:*:*:*:*:*:*:*","matchCriteriaId":"F2B146EC-2D18-430C-81B5-6F7D4328BB91"}]}]}],"references":[{"url":"https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-18.html","source":"ics-cert@hq.dhs.gov"},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01","source":"ics-cert@hq.dhs.gov","tags":["Third Party Advisory","US Government Resource"]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Third Party Advisory","US Government Resource"]}]}}]}