{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-09-29T13:04:42.788","vulnerabilities":[{"cve":{"id":"CVE-2017-5149","sourceIdentifier":"ics-cert@hq.dhs.gov","published":"2017-02-13T22:59:00.303","lastModified":"2026-06-17T01:20:03.097","vulnStatus":"Modified","cveTags":[],"descriptions":[{"lang":"en","value":"An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints."},{"lang":"es","value":"Ha sido descubierto un problema en St. Jude Medical Merlin@home, versiones anteriores a la versión 8.2.2 (modelos RF: EX1150, modelos inductivos: EX1100 y modelos inductivos: EX1100 con capacidad MerlinOnDemand). No se verifican las identidades de los puntos finales para el canal de comunicación entre el transmisor y el sitio web de St. Jude Medical, Merlin.net. Esto puede permitir que un atacante man-in-the-middle acceda o influya en las comunicaciones entre los puntos finales identificados."}],"affected":[{"source":"ics-cert@hq.dhs.gov","affectedData":[{"vendor":"n/a","product":"St. Jude Merlin@home Transmitter before 8.2.2","versions":[{"version":"St. Jude Merlin@home Transmitter before 8.2.2","status":"affected"}]}]}],"metrics":{"cvssMetricV31":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H","baseScore":8.9,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"HIGH","privilegesRequired":"NONE","userInteraction":"NONE","scope":"CHANGED","confidentialityImpact":"LOW","integrityImpact":"HIGH","availabilityImpact":"HIGH"},"exploitabilityScore":2.2,"impactScore":6.0}],"cvssMetricV2":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"2.0","vectorString":"AV:N/AC:M/Au:N/C:P/I:P/A:P","baseScore":6.8,"accessVector":"NETWORK","accessComplexity":"MEDIUM","authentication":"NONE","confidentialityImpact":"PARTIAL","integrityImpact":"PARTIAL","availabilityImpact":"PARTIAL"},"baseSeverity":"MEDIUM","exploitabilityScore":8.6,"impactScore":6.4,"acInsufInfo":false,"obtainAllPrivilege":false,"obtainUserPrivilege":false,"obtainOtherPrivilege":false,"userInteractionRequired":false}]},"weaknesses":[{"source":"nvd@nist.gov","type":"Primary","description":[{"lang":"en","value":"CWE-476"}]}],"configurations":[{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:abbott:merlin\\@home_firmware:*:*:*:*:*:*:*:*","versionEndIncluding":"8.0","matchCriteriaId":"DB01C075-4332-4FDA-94DE-64EFA3C7829D"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:abbott:merlin\\@home_ex1100:-:*:*:*:*:*:*:*","matchCriteriaId":"1D96349E-317D-4A53-81A3-1FFFDA942A36"},{"vulnerable":false,"criteria":"cpe:2.3:h:abbott:merlin\\@home_ex1150:-:*:*:*:*:*:*:*","matchCriteriaId":"E1E96A4A-6163-4EE7-A756-B751D0B0C7B9"}]}]}],"references":[{"url":"http://www.securityfocus.com/bid/95331","source":"ics-cert@hq.dhs.gov","tags":["Third Party Advisory","VDB Entry"]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01A","source":"ics-cert@hq.dhs.gov","tags":["Mitigation","Third Party Advisory","US Government Resource"]},{"url":"http://www.securityfocus.com/bid/95331","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Third Party Advisory","VDB Entry"]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01A","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Mitigation","Third Party Advisory","US Government Resource"]}]}}]}