{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-04-19T10:53:58.554","vulnerabilities":[{"cve":{"id":"CVE-2017-12722","sourceIdentifier":"ics-cert@hq.dhs.gov","published":"2018-02-15T10:29:00.383","lastModified":"2024-11-21T03:10:05.837","vulnStatus":"Modified","cveTags":[],"descriptions":[{"lang":"en","value":"An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module."},{"lang":"es","value":"Se ha descubierto un problema de lectura fuera de límites en Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, en versiones 1.1, 1.5 y 1.6. Un componente de terceros empleado en la bomba lee memoria fuera de límites, lo que provoca que el módulo de comunicaciones se cierre inesperadamente. Smiths Medical estima que el cierre inesperado del módulo de comunicaciones no tendría ningún impacto en la operación del módulo terapéutico."}],"metrics":{"cvssMetricV30":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"3.0","vectorString":"CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L","baseScore":5.3,"baseSeverity":"MEDIUM","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"NONE","integrityImpact":"NONE","availabilityImpact":"LOW"},"exploitabilityScore":3.9,"impactScore":1.4}],"cvssMetricV2":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"2.0","vectorString":"AV:N/AC:L/Au:N/C:N/I:N/A:P","baseScore":5.0,"accessVector":"NETWORK","accessComplexity":"LOW","authentication":"NONE","confidentialityImpact":"NONE","integrityImpact":"NONE","availabilityImpact":"PARTIAL"},"baseSeverity":"MEDIUM","exploitabilityScore":10.0,"impactScore":2.9,"acInsufInfo":false,"obtainAllPrivilege":false,"obtainUserPrivilege":false,"obtainOtherPrivilege":false,"userInteractionRequired":false}]},"weaknesses":[{"source":"nvd@nist.gov","type":"Primary","description":[{"lang":"en","value":"CWE-125"}]}],"configurations":[{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:*","matchCriteriaId":"E9DB2BFA-A772-4A90-B79B-D6FB2090A63C"},{"vulnerable":true,"criteria":"cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.5:*:*:*:*:*:*:*","matchCriteriaId":"AC867D0F-2FF0-4E99-87FF-84A873F56341"},{"vulnerable":true,"criteria":"cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.6:*:*:*:*:*:*:*","matchCriteriaId":"048B188C-D620-42EE-84D5-61C3F03F7FAA"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:-:*:*:*:*:*:*:*","matchCriteriaId":"9874E6FE-4025-484B-A7D4-62824A299ECA"}]}]}],"references":[{"url":"http://www.securityfocus.com/bid/100665","source":"ics-cert@hq.dhs.gov","tags":["Third Party Advisory","VDB Entry"]},{"url":"http://www.securityfocus.com/bid/101252","source":"ics-cert@hq.dhs.gov","tags":["Third Party Advisory","VDB Entry"]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A","source":"ics-cert@hq.dhs.gov","tags":["Third Party Advisory","US Government Resource"]},{"url":"http://www.securityfocus.com/bid/100665","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Third Party Advisory","VDB Entry"]},{"url":"http://www.securityfocus.com/bid/101252","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Third Party Advisory","VDB Entry"]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Third Party Advisory","US Government Resource"]}]}}]}