{"resultsPerPage":1,"startIndex":0,"totalResults":1,"format":"NVD_CVE","version":"2.0","timestamp":"2026-06-16T13:07:15.558","vulnerabilities":[{"cve":{"id":"CVE-2014-5432","sourceIdentifier":"ics-cert@hq.dhs.gov","published":"2019-03-26T16:29:00.290","lastModified":"2024-11-21T02:12:02.290","vulnStatus":"Modified","cveTags":[],"descriptions":[{"lang":"en","value":"Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes."},{"lang":"es","value":"Baxter SIGMA Spectrum Infusion System, en su versión 6.05 (modelo 35700BAX), con un módulo de batería inalámbrica (WBM), en su versión 16, es accesible de forma remota mediante el puerto 22/SSH sin autenticación. Un atacante remoto podría ser capaz de realizar cambios no autorizados en la configuración del WBM, así como lanzar comandos para acceder a las credenciales de la cuenta y a las claves compartidas. Baxter afirma que esta vulnerabilidad solo otorga acceso a características y funcionalidades del WBM y que la bomba de inyección de SIGMA Spectrum no puede ser controlada desde el WBM. Baxter ha publicado una nueva versión de SIGMA Spectrum Infusion System, la 8, que incluye cambios en el software y el hardware."}],"metrics":{"cvssMetricV30":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"3.0","vectorString":"CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","baseScore":9.8,"baseSeverity":"CRITICAL","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"HIGH","integrityImpact":"HIGH","availabilityImpact":"HIGH"},"exploitabilityScore":3.9,"impactScore":5.9}],"cvssMetricV2":[{"source":"nvd@nist.gov","type":"Primary","cvssData":{"version":"2.0","vectorString":"AV:N/AC:L/Au:N/C:P/I:P/A:P","baseScore":7.5,"accessVector":"NETWORK","accessComplexity":"LOW","authentication":"NONE","confidentialityImpact":"PARTIAL","integrityImpact":"PARTIAL","availabilityImpact":"PARTIAL"},"baseSeverity":"HIGH","exploitabilityScore":10.0,"impactScore":6.4,"acInsufInfo":false,"obtainAllPrivilege":false,"obtainUserPrivilege":false,"obtainOtherPrivilege":false,"userInteractionRequired":false}]},"weaknesses":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","description":[{"lang":"en","value":"CWE-592"}]},{"source":"nvd@nist.gov","type":"Primary","description":[{"lang":"en","value":"CWE-287"}]}],"configurations":[{"operator":"AND","nodes":[{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":true,"criteria":"cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*","matchCriteriaId":"E6B02D87-6C75-4D60-8D4C-84628757214F"}]},{"operator":"OR","negate":false,"cpeMatch":[{"vulnerable":false,"criteria":"cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*","matchCriteriaId":"49E25260-EC14-4E98-A86B-CBBE47E26AE5"},{"vulnerable":false,"criteria":"cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:*","matchCriteriaId":"175BDC61-5163-4D78-BF02-6B9EF5D38841"}]}]}],"references":[{"url":"https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01","source":"ics-cert@hq.dhs.gov","tags":["Mitigation","Third Party Advisory","US Government Resource"]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01","source":"af854a3a-2127-422b-91ae-364da2661108","tags":["Mitigation","Third Party Advisory","US Government Resource"]}]}}]}